How to address self-absorption in LIBS using millisecond time-width detectors
; Gómez Sánchez, Yenner Paul Otros responsables
Fecha
07/2025
Editorial
Pergamon-Elsevier Science Ltd
Revista
Spectrochimica Acta Part B: Atomic Spectroscopy
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0584-8547
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
To calculate self-absorption in Laser Induced Breakdown Spectroscopy (LIBS) using long-acquisition-time detectors (millisecond-range) it is necessary to account for the evolution of the plasma during the acquisition time and, consequently, the variation in the emitted spectral intensity. This study develops a model that accounts for temperature variations during the measurement period, assuming an adiabatic plasma explosion. By parameterizing all emission and absorption quantities using temperature (T), the model allows for the calculation of spectral line intensities in the presence of self-absorption. The model is applied to binary Ti–Zr alloys, where predictions of pure element spectra are compared with experimental data, and self-absorption is analyzed as stoichiometry varies in six alloys. The model’s predictions are validated against experimental curve of growth (C. O.G.) values. Furthermore, this approach provides a method to estimate photon emission probabilities (Einstein coefficients, Aki).
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa