Artículo

Mechanical evaluation of test configuration and dental implant geometry

Pazos, Leonardo Martín ; Perez, Ezequiel Martin Icon

Otros responsables

Fecha

04/2018

Editorial

IOP Publishing

Revista

Biomedical Physics and Engineering Express

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

2057-1976

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

In the present work, the mechanical behavior of a dental implant system (screw-implant plus abutment) was studied. The effect of: (i) screw-implant geometry, (ii) contact condition and (iii) specimen holder material on the mechanical performance of a dental implant system was investigated. Finite Element Analysis and mechanical tests were carried out. The results showed that, screw-implant diameter and length affected the von Mises and maximum Principal Stress distributions. These geometrical factors influence the maximal values and stress concentration zone size, respectively. Maximal displacement values were drastically diminished with increased screw-implant diameter. Simulations indicated as probable failure locations the threaded surface of screw-implant and abutment, confirmed by mechanical tests.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa