Artículo

Surface modification of zirconium by anodisation as material for permanent implants: in vitro and in vivo study

Gomez Sanchez, Andrea Valeria Icon; Ballarre, Josefina Icon; Orellano, Juan Carlos ; Duffo, Gustavo Sergio Icon; Cere, Silvia Icon

Otros responsables

Fecha

2012-09-29

Editorial

Springer

Revista

Journal of Materials Science: Materials In Medicine

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0957-4530

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Resumen

The potential use of anodised zirconium as permanent implant has been investigated. Zirconium was anodised at constant potential between 3 and 30 V in H3PO4. Electrochemical assays were conducted in simulated body fluid solution (SBF) in order to evaluate the effect of the surface oxide on the corrosion resistance in vitro after 30 days of immersion. The rupture potential increases when increasing thickness of the anodic surface film. The increase in the barrier effect when increasing anodising potential is also verified by EIS. Anodisation in H3PO4 proved to increase the apatite formation capability of zirconium in a single step. In vivo bone formation was also analysed by implanting the modified materials in Wistar rats. Anodised Zr presents higher corrosion resistance in SBF in all the studied immersion times when compared with non anodised Zr. Additionally, in vivo experiments evidence bone generation and growth in contact with zirconium implants both in the as-received and anodised condition.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa