Multivariate optimization of a solid phase extraction system employing l-tyrosine immobilized on carbon nanotubes applied to molybdenum analysis by inductively coupled plasma optical emission spectrometry with ultrasound nebulization
; Gil, Raul Andres
; Smichowski, Patricia Nora
; Pacheco, Pablo Hugo 
Otros responsables
Fecha
06/2014
Editorial
Elsevier
Revista
Microchemical Journal
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0026-265X
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
A method for Mo solid phase extraction on l-tyrosine immobilized on carbon nanotubes (l-tyr-CNTs) is presented. l-tyr-CNTs were used to fill a minicolumn and introduced into a FI system employing inductively coupled plasma optical emission spectrometry with ultrasound nebulization (USN-ICP OES). Five FI parameters such as buffer flow rate (BFR) and concentration (BC); sample flow rate (SFR); eluent flow rate (EFR) and concentration (EC); and pH were chosen for optimization employing a half fraction composite design (HFFD). Multivariate optimization through central composite design (CCD) allowed establishing the statistical ideal parameter values to reach maximum Mo signal. From HFFD and CCD it was established that SFR was not affecting the system significantly and that the optimal experimental conditions were: pH, 4.0; BC, 5 mM ammonium acetate; EC, 15% (v v− 1) and EFR, 2 mL min− 1. Under these conditions an enhancement factor of 750-fold (25 for preconcentration system and 30 for USN) was obtained reaching a detection limit of 40 ng L− 1 with a precision of 1.32%. The system was successfully applied to a certified reference material NIST CRM 1643e (trace elements in water) and river, thermal, mine and tap water samples.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa