Ultralow Power Ionizing Dose Sensor Based on Complementary Fully Depleted MOS Transistors for Radiotherapy Application
; Flandre, Denis
; André, Nicolás
; Irazoqui, Julieta ; Perez, Martín ; Gomez Berisso, Mariano
; Lipovetzky, José 
Otros responsables
Fecha
2019-10-02
Editorial
Institute of Electrical and Electronics Engineers
Revista
Ieee Transactions on Nuclear Science
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0018-9499
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
We evaluate the use of thick buried oxide (BOX) of fully depleted silicon-on-insulator (FD-SOI) transistors for total ionizing dose (TID) measurements in a radiotherapy application. The devices were fabricated by a custom process at Université Catholique de Leuven (UCL) that allows one to make accumulation mode pMOS transistors and inversion mode nMOS transistors. We characterized the temperature behavior of these devices and the response under X-ray irradiation produced by an Elekta radiotherapy linear accelerator and compared the obtained dose sensitivity with other published works. Taking advantage of these devices, an ultralow power MOS ionizing dose sensor or MOS dosimeter with inherent temperature compensation is presented. This dosimeter achieved a sensitivity of 154 mV/Gy with a temperature error factor of 13 mGy/°C and a current consumption below 1 nA.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa