Artículo

Reducing risks in type 1 diabetes using H∞ Control

Colmegna, Patricio Hernán Icon; Sanchez Peña, Ricardo Salvador Icon; Gondhalekar, Ravi ; Dassau, Eyal ; Doyle, Francis J.

Otros responsables

Fecha

12/2014

Editorial

Institute of Electrical and Electronics Engineers

Revista

Ieee Transactions On Bio-medical Engineering

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0018-9294

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Resumen

A control scheme was designed in order to reduce the risks of hyperglycemia and hypoglycemia in type 1 diabetes mellitus (T1DM). This structure is composed of three main components: an H∞ robust controller, an insulin feedback loop (IFL), and a safety mechanism (SM). A control-relevant model that is employed to design the robust controller is identified. The identification procedure is based on the distribution version of the UVA/Padova metabolic simulator using the simulation adult cohort. The SM prevents dangerous scenarios by acting upon a prediction of future glucose levels, and the IFL modifies the loop gain in order to reduce postprandial hypoglycemia risks. The procedure is tested on the complete in silico adult cohort of the UVA/Padova metabolic simulator, which has been accepted by the Food and Drug Administration (FDA) in lieu of animal trials.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa