Artículo

Noise reduction in resonator-based ultrasound sensors by using a CW laser and phase detection

Riobó, Lucas Matías Icon; Hazan, Yoav ; Veiras, Francisco Ezequiel Icon; Garea, María ; Sorichetti, Patricio Aníbal ; Rosenthal, Amir

Otros responsables

Fecha

06/2019

Editorial

Optical Society of America

Revista

Optics Letters

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0146-9592

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

The detection of ultrasound via optical resonators is conventionally performed by tuning a continuous-wave (CW) laser to the linear slope of the resonance and monitoring the intensity modulation at the resonator output. While intensity monitoring offers the advantage of simplicity, its sensitivity is often limited by the frequency noise of the CW laser. In this work, we develop an alternative CW technique that can significantly reduce measurement noise by monitoring variations in the phase, rather than intensity, at the resonator output. In our current implementation, which is based on a balanced Mach–Zehnder interferometer for phase detection, we demonstrate a 24-fold increase in the signal-to-noise ratio of the detected ultrasound signal over the conventional, intensity-monitoring approach.