Non-invasive central aortic pressure measurement: what limits its application in clinical practice?
; Zócalo, Yanina ; Salazar, Federico ; Bia, Daniel Otros responsables
Fecha
05/2023
Editorial
Frontiers Media
Revista
Frontiers in Cardiovascular Medicine
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
2297-055X
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Resumen
The following article highlights the need for methodological transparency and consensus for an accurate and non-invasive assessment of central aortic blood pressure (aoBP), which would contribute to increasing its validity and value in both clinical and physiological research settings. The recording method and site, the mathematical model used to quantify aoBP, and mainly the method applied to calibrate pulse waveforms are essential when estimating aoBP and should be considered when analyzing and/or comparing data from different works, populations and/or obtained with different approaches. Up to now, many questions remain concerning the incremental predictive ability of aoBP over peripheral blood pressure and the possible role of aoBP-guided therapy in everyday practice. In this article, we focus on “putting it on the table” and discussing the main aspects analyzed in the literature as potential determinants of the lack of consensus on the non-invasive measurement of aoBP.