Magnetic field inhomogeneities due to co2 incubator shelves: A source of experimental confounding and variability? | ||
| dc.conicet.organizacion | Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Luis. Instituto de Física Aplicada "Dr. Jorge Andrés Zgrablich". Universidad Nacional de San Luis. Facultad de Ciencias Físico Matemáticas y Naturales. Instituto de Física Aplicada "Dr. Jorge Andrés Zgrablich" | |
| dc.conicet.organizacion | Slovak Academy of Sciences | |
| dc.conicet.organizacion | Slovak Academy of Sciences | |
| dc.conicet.organizacion | Russian Academy of Sciences | |
| dc.contributor.author | Makinistian, Leonardo | |
| dc.contributor.author | Belyaev, I. | |
| dc.date.issued | 2018-02 | |
| dc.description.abstract | A thorough assessment of the static magnetic field (SMF) inside a CO2 incubator allowed us to identify non-negligible inhomogeneities close to the floor, ceiling, walls and the door. Given that incubator’s shelves are made of a non-magnetic stainless steel alloy, we did not expect any important effect of them on the SMF. Surprisingly, we did find relatively strong distortion of the SMF due to shelves. Indeed, our high-resolution maps of the SMF revealed that distortion is such that field intensities differing by a factor of up to 36 were measured on the surface of the shelf at locations only few millimetres apart from each other. Furthermore, the most intense of these fields was around five times greater than the ones found inside the incubator (without the metallic shelves in), while the lowest one was around 10 times lower, reaching the so-called hypomagnetic field range. Our findings, together with a survey of the literature on biological effects of hypomagnetic fields, soundly support the idea that SMF inhomogeneities inside incubators, especially due to shelves’ holes, are a potential source of confounding and variability in experiments with cell cultures kept in an incubator. | es |
| dc.description.fil | Fil: Makinistian, Leonardo. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Luis. Instituto de Física Aplicada "Dr. Jorge Andrés Zgrablich". Universidad Nacional de San Luis. Facultad de Ciencias Físico Matemáticas y Naturales. Instituto de Física Aplicada "Dr. Jorge Andrés Zgrablich"; Argentina. Slovak Academy of Sciences; Eslovaquia | |
| dc.description.fil | Fil: Belyaev, I.. Slovak Academy of Sciences; Eslovaquia. Russian Academy of Sciences; Rusia | |
| dc.format | application/pdf | es |
| dc.identifier.citation | Makinistian, Leonardo; Belyaev, I.; Magnetic field inhomogeneities due to co2 incubator shelves: A source of experimental confounding and variability?; The Royal Society; Royal Society Open Science; 5; 2; 2-2018; 1-10 | es |
| dc.identifier.institucional | CONICET | |
| dc.identifier.issn | 2054-5703 | es |
| dc.identifier.repo | CONICET Digital | |
| dc.identifier.uri | http://hdl.handle.net/11336/86692 | |
| dc.journal.ciudad | Londres | es |
| dc.journal.number | 2 | es |
| dc.journal.pagination | 1-10 | es |
| dc.journal.pais | Reino Unido | es |
| dc.journal.title | Royal Society Open Science | es |
| dc.journal.volume | 5 | es |
| dc.language.iso | eng | es |
| dc.publisher | The Royal Society | es |
| dc.relation.alternativeid | info:eu-repo/semantics/altIdentifier/url/https://royalsocietypublishing.org/doi/10.1098/rsos.172095 | |
| dc.relation.alternativeid | info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1098/rsos.172095 | |
| dc.rights | info:eu-repo/semantics/openAccess | * |
| dc.rights.uri | https://creativecommons.org/licenses/by/2.5/ar/ | * |
| dc.subject | BACKGROUND FIELDS OF INCUBATORS | es |
| dc.subject | BIOLOGICAL EFFECTS OF WEAK STATIC MAGNETIC FIELDS | es |
| dc.subject | STRAIN-INDUCED MAGNETIZATION OF AUSTENITIC STAINLESS STEEL | es |
| dc.subject.classification | Otras Ciencias Físicas | es |
| dc.subject.classification | Ciencias Físicas | es |
| dc.subject.classification | CIENCIAS NATURALES Y EXACTAS | es |
| dc.title | Magnetic field inhomogeneities due to co2 incubator shelves: A source of experimental confounding and variability? | es |
| dc.type | info:eu-repo/semantics/article | |
| dc.type | info:ar-repo/semantics/artículo | |
| dc.type | info:eu-repo/semantics/publishedVersion | |