Impact of Methodological and Calibration Approach on the Association of Central and Peripheral Systolic Blood Pressure with Cardiac Structure and Function in Children, Adolescents and Adults
; Bia, Daniel ; Zócalo, Yanina Otros responsables
Fecha
10/2019
Editorial
Adis
Revista
High Blood Pressure and Cardiovascular Prevention
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
1120-9879
eISSN
1179-1985
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
Introduction: Peripheral and aortic systolic blood pressure (pSBP and aoSBP) were measured using different methodological and calibration approaches to analyze the association and agreement between pSBP and/or aoSBP, and the association of pSBP and aoSBP with left ventricle (LV) and atrium (LA) structural–functional characteristics. Methods: In healthy subjects (n = 269, age: 9–85 years; n = 147, age < 24 years) LV and LA parameters were echocardiography-derived. pSBP and aoSBP were obtained by brachial sub-diastolic (Mobil-O-Graph®) and supra-systolic oscillometry (Arteriograph®) and aortic diameter waveform re-calibration (RCD; ultrasonography), using three calibration schemes: systo-diastolic (SD), calculated mean (CM), and oscillometric mean (OscM). Results: Always pSBP and aoSBP were positively associated; aoSBP obtained with the Mobil-O-Graph® and calibrated to CM or OscM were the ones that showed the lowest levels of association with the remaining forms of aoSBP and pSBP. Bland-Altman related mean errors varied noticeably (e.g. − 27, − 23, − 17, − 12 or 8 mmHg when aoSBP obtained with MOG (OscM) was compared with data from other methodological and calibration schemes). The aoSBP data obtained with Mobil-O-Graph® (calibration: CM and OscM) showed the highest levels of association with cardiac structural characteristics. aoSBP values obtained calibrating to OscM were higher than those obtained calibrating to SD or CM. Conclusions: aoSBP obtained with Mobil-O-Graph® and calibrated to CM or OscM showed (1) lower association with other forms of aoSBP and pSBP determination and (2) higher levels of association with LV and LA structural characteristics. Differences in aoSBP data between approaches were more sensitive to the calibration method than to the device used.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa