Multichannel radiometer calibration: A new approach
; Booth, Charles R. ; Armstrong, Roy ; Brunat, Claudio ; Cabrera, Sergio ; Camilion, Carolina ; Casiccia, Claudio ; Deferrari, Guillermo Alejandro
; Fuenzalida, Humberto ; Lovengreen, Charlotte ; Paladini, Alejandro Alberto
; Pedroni, Jorge ; Rosales, Alejandro ; Zagarese, Horacio Ernesto
; Vernet, Maria Otros responsables
Fecha
09/2005
Editorial
Optical Society of America
Revista
Applied Optics
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
1539-4522
eISSN
2155-3165
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
The error in irradiance measured with Sun-calibrated multichannel radiometers may be large when the solar zenith angle (SZA) increases. This could be particularly detrimental in radiometers installed at mid and high latitudes, where SZAs at noon are larger than 50° during part of the year. When a multi-regressive methodology, including the total ozone column and SZA, was applied in the calculation of the calibration constant, an important improvement was observed. By combining two different equations, an improvement was obtained at almost all the SZAs in the calibration. An independent test that compared the irradiance of a multichannel instrument and a spectroradiometer installed in Ushuaia, Argentina, was used to confirm the results.
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa