Milk fat content measurement by a simple UV spectrophotometric method: an alternative screening method
; Carmine, Mariela ; Echeverría, G. E. ; Pécora, R. P. ; Kivatinitz, Silvia Clara Otros responsables
Fecha
02/2005
Editorial
American Dairy Science Association
Revista
Journal of Dairy Science
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0022-0302
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
We developed a simple microtechnique to measure lipids in milk by UV spectrophotometry. This technique is based upon the property of fatty acids to absorb UV light proportional to their concentration. Samples of powdered or fluid milk (30 or 60 µL) are added to 3 mL of analytic grade ethanol and stored at −20°C during at least 1 hr. This procedure precipitates proteins and hydrophobic peptides that interfere with UV measurement. Sample absorbances are then measured at 208 nm in an UV-Vis spectrophotometer. This technique correlated very well against Milko-Scan , a device that measures milk fat by IR spectroscopy, with a r2 > 0.982. Accuracy and precision, evaluated by recovery and replicate assays, are also very acceptable. This method is suitable as a fast, cost-effective alternative screening method to estimate milk fat content in small samples without prior lipid extraction.