A Novel Approach to Design and Evaluate BNCT Neutron Beams Combining Physical, Radiobiological, and Dosimetric Figures of Merit

Postuma, Ian ; González, Sara Josefina Icon; Herrera, Maria Silvia Icon; Provenzano, Lucas Icon; Ferrarini, Michele ; Magni, Chiara ; Protti, Nicoletta ; Fatemi, Setareh ; Vercesi, Valerio ; Giuseppe Battistoni ; Tamburini, Umberto Anselmi ; Liu, Yuan Hao ; Kankaanranta, Leena ; Koivunoro, Hanna ; Altieri, Saverio ; Bortolussi, Silva

Otros responsables

Fecha

2021-02-26

Editorial

MDPI

Revista

Biology

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

2079-7737

eISSN

2079-7737

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

(1) Background:The quality of neutron beams for Boron Neutron Capture Therapy (BNCT) is currently defined by its physical characteristics in air. Recommendations exist to define whether a designed beam is useful for clinical treatment. This work presents a new way to evaluate neutron beams based on their clinical performance and on their safety, employing radiobiological quantities. (2) Methods: The case study is a neutron beam for deep-seated tumors from a 5 MeV proton beam coupled to a beryllium target. Physical Figures of Merit were used to design five beams; however, they did not allow a clear ranking of their quality in terms of therapeutic potential. The latter was then evaluated based on in-phantom dose distributions and on the calculation of the Uncomplicated Tumor Control Probability (UTCP). The safety of the beams was also evaluated calculating the in-patient out-of-beam dosimetry. (3) Results: All the beams ensured a UTCP comparable to the one of a clinical beam in phantom; the safety criterion allowed to choose the best candidate. When this was tested in the treatment planning of a real patient treated in Finland, the UTCP was still comparable to the one of the clinical beam. (4) Conclusions: Even when standard physical recommendations are not met, radiobiological and dosimetric criteria demonstrate to be a valid tool to select an effective and safe beam for patient treatment.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by