What characteristics a clinical CSF system has to have?
; Issolio, Luis Alberto
; Santillán, Javier Enrique
; Aguirre, Rolando Carlos 
Otros responsables
Fecha
05/2009
Editorial
Wroclaw University of Technology
Revista
Optica Applicata
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0078-5466
eISSN
1899-7015
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
We discuss the characteristics a system to measure the contrast sensitivity function (CSF) in the ophthalmologic clinic has to have. We propose that this system should be computer based in order to assure flexibility and precision. Besides the original calibration, this equipment needs an auxiliary system to keep the working conditions allowing periodical gamma corrections. We try a calibration method based on visual comparisons and show that it is valid and simple. We propose to use an adaptive psychophysical method to obtain contrast thresholds that assure a good compromise between precision and duration of the whole test. Finally, we propose that the system has to have its own normality curves for the different age ranges allowing the practitioner to perform clinical evaluations. Summing up, we can say that taking into account the above issues, the fidelity of the stimuli will be guaranteed and the challenges entailed in its transference to the clinic will be overcome.