Artículo

Application of a long-range swept source optical coherence tomography-based scheme for dimensional characterization of multilayer transparent objects

Morel, Eneas Nicolas Icon; Russo, Nelida Araceli ; Torga, Jorge Román Icon; Duchowicz, Ricardo Icon

Otros responsables

Fecha

08/2017

Editorial

Spie

Revista

Optical Engineering

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0091-3286

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

This work presents the use of a recently developed interferometric system based on the sweptsource optical coherence tomography (SS-OCT) technique, which allows the characterization of transparentand semitransparent multilayer systems employing a tunable fiber-optic laser with a coherence length suitablefor achieving long-deep range imaging (>10 cm). The inclusion of fiber Bragg gratings in the system allows it toperform a self-calibration in each sweep of the light source. Measurements carried out on cuvettes, ampoules,small bottles, and glass containers used in the pharmaceutical industry are presented. The thicknesses of thewalls and the distance between them were determined. Transparent and semitransparent objects of a multilayertype of different thicknesses were also measured. The configuration presented allows extension of the measurementrange obtainable with the usual OCT systems, demonstrating the potentiality of the proposed scheme tocarry out quality control in industrial application.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa