Exhalatory dynamic interactions between patients connected to a shared ventilation device

Garcia Eijo, Pedro Manuel ; D'adamo, Juan Gastón Leonel Icon; Bianchetti, Arturo Abel Icon; Duriez, Thomas Pierre Cornil Icon; Cabaleiro, Juan Martin Icon; Irrazabal, Célica ; Otero, Pablo ; Artana, Guillermo Osvaldo Icon

Otros responsables

Fecha

05/2021

Editorial

Public Library of Science

Revista

Plos One

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

1932-6203

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

In this work a shared pressure-controlled ventilation device for two patients is considered. By the use of different valves incorporated to the circuit, the device enables the restriction of possible cross contamination and the individualization of tidal volumes, driving pressures, and positive end expiratory pressure PEEP. Possible interactions in the expiratory dynamics of different pairs of patients are evaluated in terms of the characteristic exhalatory times. These characteristic times can not be easily established using simple linear lumped element models. For this purpose, a 1D model using the Hydraulic and Mechanical libraries in Matlab Simulink was developed. In this sense, experiments accompany this study to validate the model and characterize the different valves of the circuit. Our results show that connecting two patients in parallel to a ventilator always resulted in delays of time during the exhalation. The size of this effect depends on different parameters associated with the patients, the circuit and the ventilator. The dynamics of the exhalation of both patients is determined by the ratios between patients exhalatory resistances, compliances, driving pressures and PEEPs. Adverse effects on exhalations became less noticeable when respiratory parameters of both patients were similar, flow resistances of valves added to the circuit were negligible, and when the ventilator exhalatory valve resistance was also negligible. The asymmetries of driving pressures, compliances or resistances exacerbated the possibility of auto-PEEP and the increase in relaxation times became greater in one patient than in the other. In contrast, exhalatory dynamics were less sensitive to the ratio of PEEP imposed to the patients.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa