Technical Note: Low-Cost Sample Holder with Flooded Gasket Prevents Crevice Corrosion in Pitting Corrosion Tests
; Kappes, Mariano Alberto 
Otros responsables
Fecha
10/2022
Editorial
NACE International
Revista
Corrosion
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0010-9312
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
This work describes an inexpensive holder for corrosion experiments, developed to avoid crevice corrosion during pitting corrosion tests of corrosion-resistant alloys. It is based on a flooded gasket, incorporated in a Stern-Makrides sample holder. Unlike the Avesta cell designed by Qvarfort, the proposed sample holder does not require a pump. The sample holder attaches to prismatic or cylindrical specimens cut from bars, plates, or any other component with a thickness greater than 10 mm. The sample holder is versatile and with no limitations on chloride concentration, solution composition, or temperature. The testing solution can be deaerated or saturated with any gas if needed. The sample holding method was validated by cyclic potentiodynamic polarization experiments on Types 304, 316 stainless steel, and C-22HS nickel-based alloy. Crevice corrosion was not evident in any of the specimens tested with the proposed method.