Impact of surface treatments on the corrosion resistance of ASTM F138-F139 stainless steel for biomedical applications
; Llorente, Carlos Luis ; Bilmes, Pablo David
; Gervasi, Claudio Otros responsables
Fecha
06/2012
Editorial
Elsevier
Revista
Procedia Materials Science
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
2211-8128
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
AISI 316 LVM stainless steel type considered in ASTM F138 and F139 standards for implant devices is widely used, in particular for orthopedic surgery, because it combines good biofunctionality and acceptable biocompatibility at low costs. Adequate interaction of these materials with the human body and its capability to reach the desired service level are determined by the surface preparation. The goal of the present work is to relate the surface roughness parameters with the localized corrosion resistance of AISI 316 LVM stainless steel grit blasted for different times and passivated with nitric acid. At intermediate blasting times the roughness parameters attain an extreme value and this surface condition corresponds with the maximum pitting corrosion resistance.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-nd