Gel dosimetry measurements and Monte Carlo modeling for external radiotherapy photon beams. Comparison with a treatment planning system dose distribution

Valente, Mauro Andres Icon; Aon, Egle ; Brunetto, M. ; Castellano, Gustavo Eugenio Icon; Gallivanone, F. ; Gambarini, G.

Otros responsables

Fecha

09/2007

Editorial

Elsevier Science

Revista

Nuclear Instruments and Methods in Physics Research A: Accelerators, Spectrometers, Detectors and Associated Equipament

TĆ­tulo del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0168-9002

eISSN

ISBN

Idioma

InglƩs

Tipo del evento

Clasificación temÔtica

Resumen

Gel dosimetry has proved to be useful to determine absorbed dose distributions in radiotherapy, as well as to validate treatment plans. Gel dosimetry allows dose imaging and is particularly helpful for non-uniform dose distribution measurements, as may occur when multiple-field irradiation techniques are employed. In this work, we report gel-dosimetry measurements and Monte Carlo (PENELOPEĀ®) calculations for the dose distribution inside a tissue-equivalent phantom exposed to a typical multiple-field irradiation. Irradiations were performed with a 10 MV photon beam from a VarianĀ® Clinac 18 accelerator. The employed dosimeters consisted of layers of Fricke Xylenol Orange radiochromic gel. The method for absorbed dose imaging was based on analysis of visible light transmittance, usually detected by means of a CCD camera. With the aim of finding a simple method for light transmittance image acquisition, a commercial flatbed-like scanner was employed. The experimental and simulated dose distributions have been compared with those calculated with a commercially available treatment planning system, showing a reasonable agreement.

Colecciones

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este Ć­tem se describe como by-nc-sa