Artículo

A simple and reproducible capacitive electrode

Spinelli, Enrique Mario Icon; Guerrero, Federico Nicolás Icon; García, Pablo ; Haberman, Marcelo Alejandro Icon

Otros responsables

Fecha

03/2016

Editorial

Elsevier

Revista

Medical Engineering & Physics

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

1350-4533

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Resumen

Capacitive Electrodes (CE) allow the acquisition of biopotentials through a dielectric layer, without the use of electrolytes, just by placing them on skin or clothing, but demands front-ends with ultra-high input impedances. This must be achieved while providing a path for bias currents, calling for ultra-high value resistors and special components and construction techniques. A simple CE that uses bootstrap techniques to avoid ultra-high value components and special materials is proposed. When electrodes are placed on the skin; that is, with coupling capacitances CS of around 100 pF, they present a noise level of 3.3 μVRMS in a 0.5-100 Hz bandwidth, which is appropriate for electrocardiography (ECG) measurements. Construction details of the CE and the complete circuit, including a fast recovery feature, are presented.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa