A general tumour control probability model for non-uniform dose distributions
; Carando, Daniel Germán 
Otros responsables
Fecha
12/2008
Editorial
Oxford University Press
Revista
Mathematical Medicine And Biology-a Journal Of The Ima
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
1477-8599
eISSN
1477-8602
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Resumen
Perfectly uniform dose distributions over target volumes are almost impossible to achieve in clinical practice, due to surrounding normal tissues dose constraints that are commonly imposed to treatment plans. This article introduces a new approach to compute tumour control probabilities (TCPs) under inhomogeneous dose conditions. The equivalent subvolume model presented here does not assume independence between cell responses and can be derived from any homogeneous dose TCP model. To check the consistency of this model, some natural properties are shown to hold, including the so-called uniform dose theorem. In the spirit of the equivalent uniform dose (EUD) concept introduced by Niemierko (1997, Med. Phys., 24, 103–110), the probability-EUD is defined. This concept together with the methodology introduced to compute TCPs for inhomogeneous doses is applied to different uniform dose TCP models. As expected, the TCP takes into account the whole dose distribution over the target volume, but it is influenced more strongly by the low-dose regions. Finally, the proposed methodology and other approaches to the inhomogeneous dose scenario are compared.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa