Artículo

Detection of Volatile Organic Compound with an Ion Mobility Spectrometry Cell Type Device with a Corona Discharge Ionization Source

Boggio, Norberto Gabriel Icon; Pierpauli, Karina Alejandra ; Rinaldi, Carlos Alberto Icon; Lamagna, Alberto

Otros responsables

Fecha

04/2011

Editorial

American Scientific Publishers

Revista

Sensor Letters

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

1546-198X

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Resumen

Using the cross flow ion mobility spectrometry (CFIMS) concept, together with a pulsed corona discharge ionization source, a prototype of IMS cell type device was constructed and tested. In this work, we report the experimental conditions to set up the instrument and the analytical results based on the principal components analysis (PCA) obtained in the detection of toluene, acetone, ethanol and isoamyl acetate on gas phase. The advantages of this new design are the high ions concentration obtained compared to the commonly used 63Ni source, the simplicity of the CFIMS and the rapid response time.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa