Using Proton Nuclear Magnetic Resonance (NMR) as a calibrating reference for magnetic field measurement instruments: Sensitive volume and magnetic field homogeneity
; Forte, Guillermo Omar
; Anoardo, Esteban 
Otros responsables
Fecha
02/2020
Editorial
Elsevier
Revista
Measurement
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0263-2241
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
Nuclear magnetic resonance can be conveniently used to set up reference values of magnetic flux densities for the calibration of measurement instrumentation. Two measurement procedures are proposed based on the Fourier analysis of the nuclear magnetic signal. Particularly, we consider the situation where the reference magnetic flux density may change its value across the sensor active area/volume due to spatial inhomogeneities. An explored potential solution uses an electronic compensation system in order to minimize the spatial inhomogeneities of the magnetic flux density within the calibrating volume. For this purpose, a previously designed device was added to the magnetic resonance apparatus. Both methods allow a performance better than 10 ppm in calibrating measurements by using a magnetic flux density source of the order of 100 ppm in spatial homogeneity within the calibrating volume. Examples of both methods are discussed.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-nd