Carbon nanotubes non-covalently functionalized with cytochrome c: A new bioanalytical platform for building bienzymatic biosensors
; Venegas, Constanza J. ; Gutierrez, Alejandro
; Rivas, Gustavo Adolfo
; Bollo, Soledad Otros responsables
Fecha
09/2016
Editorial
Elsevier Science
Revista
Microchemical Journal
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0026-265X
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Resumen
We report for the first time the use of glassy carbon electrodes (GCE) modified by drop-coating with multi-walled carbon nanotubes (MWCNTs) dispersed in cytochrome c as platform for the construction of bienzymatic biosensors, using the glycoenzyme glucose oxidase (GOx) as model. The sensitivity of the resulting biosensor at -0.050 V towards glucose using hydroquinone (H2Q) as redox mediator was (96 ± 2) μA M-1, with a linear range between 1.00 × 10-4 M and 1.00 × 10-3 M and a detection limit of 8 μM. The biosensor demonstrated to be highly reproducible with RSD. values of 6.1% for five different bioelectrodes. The combination of the excellent dispersing properties of Cyt c, the stability of MWCNTs-Cyt c dispersion and the intimate interaction between Cyt c, MWCNTs and GOx, allowed the construction of an efficient glucose biosensor successfully used to determine glucose in commercial beverages.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa