ICP-OES Determination of cobalt in natural water using a flow injection system after preconcentration on activated carbon
; Gasquez, José Antonio ; Olsina, Roberto Antonio
; Martinez, Luis Dante 
Otros responsables
Fecha
11/2003
Editorial
Perkin-elmer Corp
Revista
Atomic Spectroscopy
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0195-5373
eISSN
2708-521X
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Resumen
An on-line cobalt preconcentration system is described for the ICP-OES determination of cobalt in drinking water in a flow injection system. The methodology developed is simple and rapid. Trace amounts of cobalt were pre-concentrated by sorption on a conical mini column packed with activated carbon (AC) at pH 9.5, with retention of 95%. The cobalt was removed from the conical mini column with 20% (v/v) nitric acid. An enrichment factor of 95-fold for a sample volume of 50 mL was obtained. The detection limit for the preconcentration method proposed was found to be 20 ng L–1. The precision for 10 replicate determinations at the 5-µg L–1 Co level was 2.7% relative standard deviation, calculated from the peak heights obtained. The calibration graph preconcentration method for cobalt species was linear with a correlation coefficient of 0.9994 at levels near the detection limits up to at least 100 µg L–1. The method was successfully applied to the determination of cobalt in natural water samples.