Computer-assisted surgery for replacement of the temporomandibular joint with customized prostheses: can we validate the results?
; Ritacco, Lucas 
Otros responsables
Fecha
09/2020
Editorial
Springer
Revista
Oral and Maxillofacial Surgery
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
1865-1550
eISSN
1865-1569
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
Purpose: Replacing the temporomandibular joint poses an important challenge to maxillofacial surgeons, and for certain disorders, it represents the treatment’s gold standard. Computer-assisted surgery (comprising preoperative virtual planning, virtual intraoperative navigation and 3D printing) is a useful tool for this type of surgery. However, we do not know if and how much the final position of the prosthesis differs, in absolute values, from what was planned virtually in the preoperative phase. We propose a comparative result validation system for temporomandibular joint replacement Methods: In the present study, we propose a comparative validation system using overlapping images, between the model obtained with preoperative virtual planning and the postoperative result. Results: The mean difference for all screws of the glenoid prosthesis was 2.08 mm (range, 1.20–3.03) and for all screws of the condylar prosthesis it was 2.33 mm (range, 1.16–3.56). Mean overall difference between both prostheses in all patients was 2.21 mm (range, 1.16–3.56). Conclusions: The validation system proposed by overlapping pre- and postoperative images in temporomandibular joint replacement allowed us to establish differences in absolute values between the virtual preoperative model and the actual postoperative result expressed in millimeters.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa