Determination of the total and extractable fraction of Ni in lake sediments and natural waters of San Luis (Argentina) by FAAS using a simple solid phase extraction system
; Blanchet, A. J. ; Sombra, Lorena Luján
; Salonia, José Antonio ; Gasquez, José Antonio Otros responsables
Fecha
04/2014
Editorial
Elsevier
Revista
Microchemical Journal
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0026-265X
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
An on-line nickel preconcentration and determination system implemented with flame atomic absorption spectrometry (FAAS) was developed. The nickel was retained on the inner walls of a column filled with ethyl vinyl acetate (EVA). The EVA was used as sorbent material for on-line solid-phase extraction (SPE) in the absence of complexing agent. The effects of sampling volume, elution conditions, sample pH and flow rate on the Ni extraction were studied. The analytical efficiency parameters obtained with this methodology were: enrichment factor (EF) of 46 times for a preconcentration time of 120 s, the limit of detection (LOD) was 3.78 μg L− 1, and the precision for six replicate determinations (repeatability conditions) at 20 μg L− 1 Ni concentration was 4.38% relative standard deviation. The sampling frequency for the proposed system was about 20 samples/h. The optimized procedure was applied to the determination of Ni in natural water and lake sediment samples.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-nd