Switched LPV Glucose Control in Type 1 Diabetes
; Sánchez Peña, Ricardo ; Gondhalekar, Ravi ; Dassau, Eyal ; Doyle, Frank J. Otros responsables
Fecha
06/2016
Editorial
Institute of Electrical and Electronics Engineers
Revista
Ieee Transactions On Bio-medical Engineering
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0018-9294
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
Objective: The purpose of this paper is to regulate the blood glucose level in Type 1 Diabetes Mellitus patients with a practical and flexible procedure that can switch among a finite number of distinct controllers, depending on the user's choice. Methods: A switched linear parameter-varying controller with multiple switching regions, related to hypo-, hyper-, and euglycemia situations, is designed. The key feature is to arrange the controller into a framework that provides stability and performance guaranty. Results: The closed-loop performance is tested on the complete in silico adult cohort of the UVA/Padova metabolic simulator, which has been accepted by the U.S. Food and Drug Administration in lieu of animal trials. The outcome produces comparable or improved results with respect to previous works. Conclusion: The strategy is practical because it is based on a model tuned only with a priori patient information in order to cover the interpatient uncertainty. Results confirm that this control structure yields tangible improvements in minimizing risks of hyper-and hypoglycemia in scenarios with unannounced meals. Significance: This flexible procedure opens the possibility of taking into account, at the design stage, unannounced meals and/or patients' physical exercise.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa