Optimized EDXRF system for simultaneous detection of gold and silver nanoparticles in tumor phantom

Santibáñez, M. ; Saavedra, R. ; Vedelago, José Alberto Icon; Malano, Francisco Mauricio Icon; Valente, Mauro Andres Icon

Otros responsables

Fecha

2019-12-24

Editorial

Pergamon-Elsevier Science Ltd

Revista

Radiation Physics and Chemistry (Oxford)

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0969-806X

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Resumen

An energy-dispersive X-ray fluorescence system (EDXRF) was optimized for simultaneous detection of gold and silver nanoparticles inside water-equivalent phantoms applied to theranostics. The optimization process was carried out in order to maximize the fluorescence detection, keeping dose levels as low as possible. Gold and silver fluorescent emissions were simultaneously detected emerging from phantom's volume, allowing the future development of multi-parametric imaging associated to specific tumor characteristics. For fluorescence detection, gold LIII-edge and silver K-edge emission lines were used, exited with a conventional X-ray source with a tungsten target. Several combinations of filters of different thickness were used to hardening the Bremsstrahlung spectrum, thus producing energetically narrow beams with central energy according to both excitation edges. In the case of gold, incident spectrum was optimized by means of strontium (Sr) and yttrium (Y) filters to relatively incrementing photons capable of exciting gold LIII-edge, while silver K-edge excitation was improved using a tin (Sn) filter. Filtering combinations made of 212.0 μm Sr with 94.5 μm Sn, and 130.2 μm Y with 94.5 μm Sn maximized fluorescence detection sensitivity and minimized delivered dose, resulting in doses 2.24 and 2.36 times lower than individual gold or silver fluorescent detection. Furthermore, when dual gold and silver fluorescent detection was performed, the minimum detectable concentration was lower than single element detection: 0.042 ± 0.002 mg/mL of Au and 0.024 ± 0.005 mg/mL of Ag in dual detection against 0.058 ± 0.003 mg/mL of Au and 0.124 ± 0.007 mg/mL of Ag when measured independently.

Colecciones

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-nd