Evaluation of an automated chemiluminescent immunoassay for salivary cortisol measurement. Utility in the diagnosis of Cushing's syndrome

Repetto, Esteban Martín Icon; Gonzalez, Diego Javier ; Jacobsen, Dario ; Smithuis, Fernando ; Jamardo, Juan ; Cano, Mariel ; Aranda, Claudio Icon; Oneto, Adriana ; Berg, Gabriela Alicia Icon; Fabre, Bibiana

Otros responsables

Fecha

03/2017

Editorial

De Gruyter

Revista

Clinical Chemistry and Laboratory Medicine

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

1434-6621

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

The aim of the current study was to validate an automated chemiluminescent immunoassay for cortisol quantification in saliva and to determine the cut-off level for Cushing´s Syndrome diagnosis in an adult population. The salivary cortisol levels from healthy individuals and patients previously diagnosed with Cushing´s Syndrome were measured with Siemens Immulite 2000 ® automatic chemiluminescent enzyme-immunoassay analyser with a brief modification. The limit of quantification was 3,4 nmol/L. The assay also demonstrated a high level of accuracy, as determined by linearity under dilution and recovery tests. Using this methodology, levels higher than 5.7 nmol/L is indicative of Cushing´s Syndrome with a diagnostic sensitivity and specificity of 90,5% and 97.4% respectively. The assay validated in the present study could be used in the clinical laboratories for the evaluation of cortisol levels in patients in which Cushing´s Syndrome is suspected.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa