Artículo

On-line preconcentration and analysis of zinc in water by flow injection-knotted reactor: Application to geological and environmental exploration

Salonia, José Antonio ; Cerutti, Estela Soledad Icon; Martinez, Luis Dante Icon; Fernandez-turiel, Jose Luis ; Gasquez, José Antonio

Otros responsables

Fecha

05/2008

Editorial

Taylor & Francis

Revista

Instrumentation Science And Technology

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

1073-9149

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

A preconcentration on-line system composed of flow injection (FI) and a knotted reactor (KR), associated with inductively coupled plasma atomic emission spectrometry (ICP-AES) has been used to determine Zn concentrations in river water. The zinc was retained as Zn-2(5-bromo-2-pyridilazo)5-diethylaminofenol precipitated complex on the knotted reactor (KR). The limit of detection (0.09gL-1 of Zn) of this method makes it particularly suitable for hydrochemical exploration in geological and environmental projects. This application has been validated in the volcanogenic gold (copper, zinc, lead) deposit of La Carolina in the province of San Luis, Argentina. The Carolina Stream and the Grande River have been sampled for water downstream this deposit. The regional background in river water is 2gL-1. Six times this concentration has been observed at more than 6.5km from the source, La Carolina mine.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa