Artículo

Liquid chromatography coupled to molecular fluorescence with postcolumn UV sensitization for thimerosal and derivative compounds monitoring in environmental samples

Acosta, Maria Gimena Icon; Torres, Alberto Sabier Icon; Kaplan, Marcos Manuel Icon; Fernandez, Liliana Patricia Icon; Pacheco, Pablo Hugo Icon; Gil, Raul Andres Icon

Otros responsables

Fecha

10/2016

Editorial

Wiley VCH Verlag

Revista

Electrophoresis

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0173-0835

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

A HPLC coupled with molecular fluorescence (MF) spectrometry method for determination of thimerosal (THM, sodium ethylmercurythiosalicylate, C9H9HgNaO2S), and derivatives is proposed. A sensitization of MF was provoked by UV irradiation of analytes in a home-made photoreactor that served as interface between the LC column and MF spectrometer. This method is applied to determination of THM, ethyl mercury, and thiosalicylic acid in samples of pharmaceutical industry effluents, and waters of La Carolina and Jáchal rivers situated in the center-west side of San Luis city and in the east of San Juan city (Middle West, Argentine) where the effluents are dumped. The LODs calculated on basis of 3σ criterion were 1.8, 5, and 0.05 μmol/L for THM, ethyl mercury, and for thiosalicylic acid, respectively.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa