Hydrophilic interaction liquid chromatography–tandem mass spectrometry (HILIC-MS/MS) determination of cocaine and its metabolites benzoylecgonine, ecgonine methyl ester, and cocaethylene in hair samples
; Jurado, Carmen ; López-Rivadulla, Manuel Otros responsables
Fecha
01/2010
Editorial
Springer Heidelberg
Revista
Analytical and Bioanalytical Chemistry
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
1618-2642
eISSN
1618-2650
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
This study reports the development and validation of a method using hydrophilic interaction liquid chromatography-tandem mass spectrometry (HILIC-MS/ MS) for the analysis of cocaine and its metabolites benzoylecgonine (BE), ecgonine methyl ester (EME), and cocaethylene (CE) in hair samples. Decontamination was performed as follows: Firstly, the aliquot of hair was briefly rinsed with 2 mL dichloromethane, then was washed three times with 10 mL 0.01 M phosphate buffer, pH 6, for 15 min, followed by 2 mL 2-propanol for less than 2 min, and, finally, a last rinse with 2 mL dichloromethane was again done. Cocaine compounds were extracted from 10 mg of hair by incubation with 2 mL 0.1 M HCl at 50 °C for 12 h and purified by solid phase extraction with Oasis MCX cartridges. Analysis was performed by LC-MS/ MS using an Atlantis HILIC silica chromatographic column. The method was fully validated. Linearity was established over the concentration range 0.020?10.0 ng/mg for cocaine (COC), 0.010?10.0 ng/mg for BE and CE, and 0.005?2.0 ng/mg for EME, and the correlation coefficients were all >0.99. Extraction efficiency was >70% for all analytes. Limits of detection were 0.0005 ng/mg for CE and 0.001 ng/mg for the other analytes (COC, BE, and EME). Lower limits of quantification were the lowest points of the calibration curves with acceptable accuracy and precision (coefficient of variation ≤20%). Intra- and inter-day imprecision ranged between 1.5% and 9.5% and 0.7% and 12.6%, respectively. Intra- and inter-day inaccuracy ranged from 0.5% to 12.3% and from 0.7% to 7.1%, respectively. With regard to matrix effects, suppression was <27.5% in all cases. The method was applied to the analysis of several samples derived from forensic cases.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa