Miniaturized HPLC-UV method for analysis of Coenzyme Q10 in human plasma
; Lucangioli, Silvia Edith
; Martinefski, Manuela Romina
; Flor, Sabrina Andrea
; Tripodi, Valeria Paula 
Otros responsables
Fecha
04/2011
Editorial
Taylor & Francis
Revista
Journal Of Liquid Chromatography And Related Technologies
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
1082-6076
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
Coenzyme Q10 (CoQ10) is a cofactor in the respiratory chain and a potent endogenous antioxidant. Abnormally low levels of CoQ10 in the circulatory system are often involved with pathological states such as heart failure, mitochondrial and muscular diseases, and cancer. The development of simple, rapid, and highly sensitive methods capable of quantifying low CoQ10 levels in plasma is, therefore, increasingly required.
In this work, we developed a miniaturized HPLC-UV system for the determination CoQ10 in human plasma using an XTerra microcolumn (50 mm × 2.1 mm i.d., 3.5 µm particle size), methanol:water (98:2, v/v) as mobile phase, 275 nm, flow rate of 0.3 mL/min, 30 °C column temperature, and 2 µL injection volume. The extraction procedure consists of a plasma precipitation with 1-propanol and evaporation under nitrogen to allow a 1.5-fold enrichment. The chromatographic analysis was accomplished in 7 min. As a result, it was possible to quantitate CoQ10 in small sample volumes down to 0.07 µM and detect as low as 0.02 µM in real plasma, with good accuracy and precision, even in pathological conditions. The proposed method was found to be suitable for routine CoQ10 determination in clinical laboratories.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa