Artículo

Floating Gate sensor for in-vivo dosimetry in radiation therapies: design and first characterization

Faigon, Adrián Néstor Icon; Martinez Vazquez, Ignacio José ; Carbonetto, Sebastián Horacio Icon; García Inza, Mariano Andrés Icon

Otros responsables

Fecha

03/2017

Editorial

IOP Publishing

Revista

Journal of Physics: Conference Series

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

1742-6588

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

A floating gate dosimeter was designed and fabricated in a standard CMOS technology. The design guides and characterization are presented. The characterization included the controlled charging by tunneling of the floating gate, and its discharging under irradiation while measuring the transistor drain current whose change is the measure of the absorbed dose. The resolution of the obtained device is close to 1 cGy satisfying the requirements for most radiation therapies dosimetry. Pending statistical proofs, the dosimeter is a potential candidate for wide in-vivo control of radiotherapy treatments.