The densities of mixtures of ordinary water and 18O enriched water have been measured all over the range of compositions, at temperatures between 293.15 and 313.15 K, using a vibrating-tube densitometer. From the error analysis of the density data in the 18O enriched water region (atom % 18O > 95) we concluded that densitometry measurements with an accuracy of ±0.00002 g cm− 3 , provide a simple, precise and low-cost method to determine the 18O enrichment degree of recycled heavy oxygen water used to generate radiopharmaceuticals, such as [ 18F]fluoro-2-deoxy-D-glucose for the Positron-Emission Tomography (PET) scanning.