Biocompatibility of pure iron for biodegradable stents: Effects on human coronary aorta endothelial cells
; Ceré, Silvia 
Otros responsables
Fecha
07/2024
Editorial
Materials Research Society
Revista
Journal Of Materials Research
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0884-2914
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
Iron (Fe)-based metals are promising for the development of biodegradable cardiovascular stents due to their mechanical strength, ductility, and degradability. However, Fe corrosion generates compounds that can damage cells, but this remains insufciently explored. This study investigated the efects of pure iron on human coronary artery endothelial cells through a series of in vitro assays. Surface morphology, roughness, hydrophilicity, and chemical composition of iron—including a pre-incubated condition that mimics a 24 h implantation period—were characterized and related with cellular responses. Overall results indicated that cellular morphology and proliferation were negatively afected by the direct contact with iron but not by it extracts or indirect contact, suggesting a role of short-half-life molecules and insoluble products of iron corrosion. Interestingly, iron pre-incubated with protein-supplemented serum enhanced cellular angiogenic function.