Enhanced application of square wave voltammetry with glassy carbon electrode coupled to multivariate calibration tools for the determination of B6 and B12 vitamins in pharmaceutical preparations
; Goicoechea, Hector Casimiro 
Otros responsables
Fecha
05/2003
Editorial
Elsevier Science
Revista
Talanta
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0039-9140
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
Vitamins B6 (VB6) and B12 (VB12) were simultaneously determined in pharmaceutical preparations by using square wave voltametry (SWV) together with artificial neural networks (ANNs). Supporting electrolyte solution, pH and voltametric technique were optimised. The calibration set was built with several artificial samples containing both active ingredients and excipients. Deviations from linearity were observed for both analytes. It is probably due to interactions among the electro active components and competition by the electrode surface, fact that supports the use of ANNs. Recoveries when analysing a nine sample validation set, of 100.2 and 96.4 were calculated for VB6 and VB12, respectively. Commercial samples were analysed with reasonably good results considering the complexity of the mixture studied.