Artículo

Standardization of Ga-68 by coincidence measurements, liquid scintillation counting and 4πγ counting

Roteta, Miguel ; Peyres, Virginia ; Rodríguez Barquero, Leonor ; García Toraño, Eduardo ; Arenillas, Pablo ; Balpardo, Christian Guillermo ; Rodrigues Ferreira Maltez, Dario Pablo Icon; Llovera, Roberto

Otros responsables

Fecha

09/2012

Editorial

Pergamon-Elsevier Science Ltd

Revista

Applied Radiation and Isotopes

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0969-8043

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

The radionuclide 68Ga is one of the few positron emitters that can be prepared in-house without the use of a cyclotron. It disintegrates to the ground state of 68Zn partially by positron emission (89.1%) with a maximum energy of 1899.1 keV, and partially by electron capture (10.9%). This nuclide has been standardized in the frame of a cooperation project between the Radionuclide Metrology laboratories from CIEMAT (Spain) and CNEA (Argentina). Measurements involved several techniques: 4pbg coincidences, integral gamma counting and Liquid Scintillation Counting using the triple to double coincidence ratio and the CIEMAT/NIST methods. Given the short half-life of the radionuclide assayed, a direct comparison between results from both laboratories was excluded and a comparison of experimental efficiencies of similar NaI detectors was used instead.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa