Very sensitive electrochemical sensor for moniliformin detectionin maize samples

Diaz Toro, Paulo Cesar Icon; Arevalo, Fernando Javier Icon; Fumero, María Verónica Icon; Zon, María Alicia Icon; Fernandez, Hector Icon

Otros responsables

Fecha

03/2016

Editorial

Elsevier Science SA

Revista

Sensors and Actuators B: Chemical

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0925-4005

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Resumen

A highly sensitive electrochemical sensor based on self-assembled monolayers (SAMs) of cysteamine on gold electrodes (Au-CA) in pH 4 citrate buffer solutions was developed to quantify the moniliformin (MON) mycotoxin by cyclic voltammetry in contaminated maize samples. Parameters of SAMs modified gold electrodes such as concentration of thiol, modification time and MON accumulation time were adjusted to obtain the best performance of the sensor. The calibration curve was linear in the MON concentration range from 1 × 10-9 to 1 × 10-7 mol L-1. A limit of detection of 8.3 × 10-10 mol L-1 (0.1 ppb) was obtained. The standard addition method was used to minimize matrix effects. The concentration of MON determined with the electrochemical sensor was in very good agreement with results found by HPLC UV-vis. The sensor showed a good analytical performance with a relative standard deviation (RDS%) of about 3%.