Spectrophotometric determination of dithizone-mercury complex by solid phase microextraction in micropipette tip syringe packed with activated carbon xerogel
; Santiago, Liliana Gabriela ; Vega, Mabel del Carmen Otros responsables
Fecha
11/2016
Editorial
Elsevier Science
Revista
Microchemical Journal
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0026-265X
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
A new simple miniaturized solid phase microextraction system in a syringe was developed for mercury (Hg) preconcentration of in water samples based on the adsorption of its dithizone complex on a microcolumn. The syringe system contains a micropipette tip packed with activated carbon xerogel. The influence of acidity, eluting agents, amount of adsorbent, complex stability and interfering ions was studied in detail. The complex retained on activated carbon xerogel in the micropipette tip-syringe system could be eluted using ethanol and the concentration was determined by visible spectrophotometry.The method showed good linearity in the range of 30-280 μg L-1, with a limit of detection of 7.6 μg L-1. The relative standard deviation (RSD%) for six replicate measurements of 280 μg L-1 of Hg2+ ion was 1.36%.The proposed method was successfully applied to quantitative determination of Hg2+ in ground water and mineral water with satisfactory results.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-nd