Artículo

Spectrophotometric determination of dithizone-mercury complex by solid phase microextraction in micropipette tip syringe packed with activated carbon xerogel

Carrera Baquero, Sebastián Eduardo Icon; Santiago, Liliana Gabriela ; Vega, Mabel del Carmen

Otros responsables

Fecha

11/2016

Editorial

Elsevier Science

Revista

Microchemical Journal

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0026-265X

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

A new simple miniaturized solid phase microextraction system in a syringe was developed for mercury (Hg) preconcentration of in water samples based on the adsorption of its dithizone complex on a microcolumn. The syringe system contains a micropipette tip packed with activated carbon xerogel. The influence of acidity, eluting agents, amount of adsorbent, complex stability and interfering ions was studied in detail. The complex retained on activated carbon xerogel in the micropipette tip-syringe system could be eluted using ethanol and the concentration was determined by visible spectrophotometry.The method showed good linearity in the range of 30-280 μg L-1, with a limit of detection of 7.6 μg L-1. The relative standard deviation (RSD%) for six replicate measurements of 280 μg L-1 of Hg2+ ion was 1.36%.The proposed method was successfully applied to quantitative determination of Hg2+ in ground water and mineral water with satisfactory results.