Artículo

High gain driven right leg circuit for dry electrode systems

Guerrero, Federico Nicolás Icon; Spinelli, Enrique Mario Icon

Otros responsables

Fecha

01/2017

Editorial

Elsevier

Revista

Medical Engineering & Physics

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

1350-4533

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Resumen

This paper presents an improved driven right leg (DRL) circuit compensation together with a practical implementation. The proposed design allows to increase common mode voltage attenuation compared with the widely used dominant pole compensation while maintaining the same proven stability margin and design criteria, and requiring only a modification of its passive feedback network. A sample implementation of the proposed DRL was obtained estimating the values of interference model parameters for a dry electrode measurement system. A dominant pole compensated DRL with the same stability margin was also implemented in order to experimentally validate the proposed design against this established alternative. Measurements were conducted under both controlled and uncontrolled interference conditions. The proposed compensation experimentally demonstrated achieving a better reduction of power line harmonics, with a peak comparative improvement of around 18  dB at 50  Hz.

Colecciones

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa