Photografted methacrylate-based monolithic columns coated with cellulose tris(3,5-dimethylphenylcarbamate) for chiral separation in CEC
; Carrasco Correa, Enrique Javier ; Keunchkarian, Sonia
; Reta, Mario Roberto
; Herrero Martinez, José Manuel Otros responsables
Fecha
03/2018
Editorial
Wiley VCH Verlag
Revista
Journal of Separation Science
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
1615-9306
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Resumen
A chiral capillary monolithic column for enantiomer separation in capillary electrochromatography was prepared by coating cellulose tris(3,5-dimethylphenylcarbamate) on porous glycidyl methacrylate-co-ethylene dimethacrylate monolith in capillary format grafted with chains of [2(methacryloyloxy)ethyl] trimethylammonium chloride. The surface modification of the monolith by the photografting of [2(methacryloyloxy)ethyl] trimethylammonium chloride monomer as well as the coating conditions of cellulose tris(3,5-dimethylphenylcarbamate) onto the grafted monolithic scaffold were optimized to obtain a stable and reproducible chiral stationary phase for capillary electrochromatography. The effect of organic modifier (acetonitrile) in aqueous mobile phase for the enantiomer separation by capillary electrochromatography was also investigated. Several pairs of enantiomers including acidic, neutral, and basic analytes were tested and most of them were partially or completely resolved under aqueous mobile phases. The prepared monolithic chiral stationary phases exhibited a good stability, repeatability, and column-to-column reproducibility, with relative standard deviations below 11% in the studied electrochromatographic parameters.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by