Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings

Valverde, Esteban Raul Icon; Arini, Pedro David Icon; Biagetti, Marcelo ; Bertran, Guillermo Claudio Icon; Quinteiro, Ricardo Alberto

Otros responsables

Fecha

12/2004

Editorial

Taylor & Francis

Revista

Journal Of Medical Engineering And Technology

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0309-1902

eISSN

1464-522X

ISBN

Idioma

Inglés

Tipo del evento

Resumen

We analyzed the effects of the electrode impedance on the transfer response of a one-stage improved buffer amplifier. The electrode DC resistance (Rd) modifies the one-stage buffer transfer response. We found a limit electrode resistance (Rd(lim)) which depends on the transfer damping factor (∈). If Rd is lower than 86.5 kW, the transfer response of the buffer fulfils the American Heart Association, AHA, recommendations, but when Rd is greater than Rd(lim) it must be cautiously weighted-up because its influence in the transfer response becomes appreciable. The maximum Rd that can be driven by the buffer is 1.2 MW. Higher values never fulfil the AHA recommendations. Therefore, electrodes with higher impedance should not be used with this kind of buffer. On the other hand, when this buffer is used to build-in an instrumentation amplifier (IA) for bipolar recording, the common-mode rejection ratio (CMRR) is sensitive to the electrode type used.