Artículo

Gait Event Detection on Stair Walking Using a Rate Gyroscope

Catalfamo Formento, Paola Andrea Lucia Icon; Acevedo, Rubén Carlos ; Ghoussayni, Salim ; Ewins, David

Otros responsables

Fecha

03/2014

Editorial

Molecular Diversity Preservation International

Revista

Sensors

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

1424-8220

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

Gyroscopes have been proposed as sensors for ambulatory gait analysis and functional electrical stimulation systems. These applications often require detection of the initial contact (IC) of the foot with the floor and/or final contact or foot off (FO) from the floor during outdoor walking. Previous investigations have reported the use of a single gyroscope placed on the shank for detection of IC and FO on level ground and incline walking. This paper describes the evaluation of a gyroscope placed on the shank for determination of IC and FO in subjects ascending and descending a set of stairs. Performance was compared with a reference pressure measurement system. The absolute mean difference between the gyroscope and the reference was less than 45 ms for IC and better than 135 ms for FO for both activities. Detection success was over 93%. These results provide preliminary evidence supporting the use of a gyroscope for gait event detection when walking up and down stairs.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by