A new algorithm for estimating the rod volume fraction and the trabecular thickness from in vivo computed tomography
; Peña, Jaime Andrés ; Lu, Yongtao ; Huber, Gerd ; Morlock, Michael ; Glüer, Claus Christian ; Delrieux, Claudio Augusto 
Otros responsables
Fecha
12/2016
Editorial
American Association of Physicists in Medicine
Revista
Medical Physics
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
0094-2405
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Palabras clave
Resumen
Purpose: Existing microstructure parameters are able to predict vertebral in vitro failure load, but for noisy in vivo data more complex algorithms are needed for a robust assessment. Methods: A new algorithm is proposed for the microstructural analysis of trabecular bone under in vivo quantitative computed tomography (QCT). Five fractal parameters are computed: (1) the average local fractal dimension FD, (2) its standard deviation FD.SD, (3) the fractal rod volume ratio fRV/BV, (4) the average fractal trabecular thickness fTb.Th, and (5) its coefcient of variation fTb.Th.CV. The algorithm requires neither an explicit skeletonization of the trabecular bone, nor a well-defned transition between bone and marrow phases. Two experiments were conducted to compare the fractal with established microstructural parameters. In the frst, 20 volumes-of-interest of embedded vertebrae phantoms were scanned fve times under QCT and high-resolution (HR-)QCT and once under peripheral HRQCT (HRpQCT), to derive accuracy and precision. In the second experiment, correlations between in vitro HRQCT structural parameters were obtained from 76 human T11, T12, or L1 vertebrae. In vitro fracture data were available for a subset of 17 human T12 vertebrae so that linear regression models between failure load and microstructural HRQCT parameters could be analyzed. Results: The results showed correlations of fTb.Th and fRV/BV with their nonfractal pendants trabecular thickness (Tb.Th) and respective structure model index (SMI) while higher precision and accuracy was observed on the fractal measures. Linear models of bone mineral density with two and three fractal microstructural HRQCT parameters explained 86% and 90% (adjusted R2) of the failure load and signifcantly improved the linear models based only on BMD and established standard microstructural parameters (68%-77% adjusted R2). Conclusions: The application of fractal methods may grant further insight into the study of bone quality in vivo when image resolution and quality are less than optimal for current standard methods.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa