Simulation of intra-saccular devices for pre-operative device size selection: Method and validation for sizing and porosity simulation

Muñoz, Romina Luciana Icon; Narata, Ana Paula ; Obradó, Laura ; Hernández, Héctor ; Larrabide, Ignacio Icon

Otros responsables

Fecha

08/2022

Editorial

Pergamon-Elsevier Science Ltd

Revista

Computers In Biology And Medicine

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0010-4825

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

Intra-saccular devices (ID) are novel braided devices used for complex intracranial aneurysms treatment. Treatment success is associated with correct device size selection. A technique that predicts the ID size within the aneurysm before intervention will provide a powerful computational tool to aid the interventionist during device selection. We present a method to calculate the device's final height, radial expansion and porosity within the patient's anatomy, which allows assessing different device sizes before treatment takes place. The proposed sizing technique was tested in-vitro and in real patient's geometries obtained from 3DRA angiographic images of 8 unruptured aneurysms previously treated with IDs. The obtained simulated height was compared to the real height, with a mean error of less than 0.28 mm (±0.44). The porosity calculation method was tested in-vitro with an error of 0.02 (±0.022). The results of both sizing and porosity experiments resemble well measures from real patients. This methodology could be used before treatment to provide the interventionist with additional information that allows selecting the device that best fits the patient's aneurysm to be treated.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa