Comparison of two MALDI-TOF MS systems for the identification of clinically relevant anaerobic bacteria in Argentina

Litterio, Mirta ; Castello, Liliana ; Venuta, María Elena ; Abel, Sofía ; Fernández Canigia, Liliana Icon; Legaria, María Cristina ; Rollet, Raquel ; Vaustat, Claudia Daniela ; Azula, Natalia ; Fox, Bárbara ; Otero, Silvina ; Maldonado, María Laura ; Mangieri, Natalia Alejandra ; Rossetti, María Adelaida ; Predari, Silvia Carla ; Cejas, Daniela Icon; Barberis, Claudia

Otros responsables

Fecha

02/2024

Editorial

Asociación Argentina de Microbiología

Revista

Revista Argentina de Microbiología

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0325-7541

eISSN

1851-7617

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

The aim of this study was to compare the performance of two MALDI-TOF MS systems in the identification of clinically relevant strict anaerobic bacteria. The 16S rRNA gene sequencing was the gold standard method when discrepancies or inconsistencies were observed between platforms. A total of 333 isolates were recovered from clinical samples of different centers in Buenos Aires City between 2016 and 2021. The isolates were identified in duplicate using two MALDI-TOF MS systems, BD Bruker Biotyper (Bruker Daltonics, Bremen, Germany) and Vitek MS (bioMèrieux, Marcy-l´Etoile, France). Using the Vitek MS system, the identification of anaerobic isolates yielded the following percentages: 65.5% (n: 218) at the species or species-complex level, 71.2% (n: 237) at the genus level, 29.4% (n: 98) with no identification and 5.1% (n: 17) with misidentification. Using the Bruker Biotyper system, the identification rates were as follows: 85.3% (n: 284) at the species or species-complex level, 89.7% (n: 299) at the genus level, 14.1% (n: 47) with no identification and 0.6% (n: 2) with misidentification. Differences in the performance of both methods were statistically significant (p-values <0.0001). In conclusion, MALDI-TOF MS systems speed up microbial identification and are particularly effective for slow-growing microorganisms, such as anaerobic bacteria, which are difficult to identify by traditional methods. In this study, the Bruker system showed greater accuracy than the Vitek system. In order to be truly effective, it is essential to update the databases of both systems by increasing the number of each main spectrum profile within the platforms.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa