Skin-electrode impedance measurement during ECG acquisition: Method's validation
; la Mura, Guillermo Mario Otros responsables
Fecha
04/2016
Editorial
IOP Publishing
Revista
Journal of Physics: Conference Series
Título del libro
Nombre del evento
Fecha del evento
Institución organizadora
ISSN
1742-6596
eISSN
ISBN
Idioma
Inglés
Tipo del evento
Clasificación temática
Resumen
Skm-electrode impedance measurement can provide valuable information prior. dunng and post electrocardiographic (ECG) or electroencephalographs (EEG) acquisitions. In this work we validate a method for skm-electrode impedance measurement using test circuits with known resistance and capacitor values, at different frequencies for injected excitation current. Finally the method is successfully used for impedance measurement during ECG acquisition on a subject usmg 125 Hz and 6 nA square wave excitation signal at instrumentation amplifier mput. The method can be used for many electrodes configuration.
Colecciones
Aprobación
Revisión
Complementado por
Referenciado por
Licencia Creative Commons
Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa