Determination of lead in drinking water by ICP-AES with ultrasonic nebulization and flow-injection on-line preconcentration using an amberlite XAD-16 resin

Wuilloud, Rodolfo German Icon; Acevedo, H.A. ; Vazquez, F.A. ; Martinez, Luis Dante Icon

Otros responsables

Fecha

10/2002

Editorial

Taylor & Francis

Revista

Analytical Letters

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0003-2719

eISSN

ISBN

Idioma

Inglés

Tipo del evento

Clasificación temática

Resumen

An on-line lead preconcentration and determination system implemented with inductively coupled plasma atomic emission spectrometry (ICP-AES) associated to flow injection (FI) with ultrasonic nebulization (USN) was studied. For the retention of lead, 2-(5-bromo-2-pyridylazo)-5-diethyl-aminophenol (5-Br-PADAP) and Amberlite XAD-16 were used, at pH 9.2. The lead was removed from the microcolumn in counter flow with nitric acid. A total enhancement factor of 150 was obtained with respect to ICP-AES using pneumatic nebulization (15.0 for USN and 10.0 for column). The detection limit value for the preconcentration of 10 mL of aqueous solution of Pb was 0.2 ng mL-1. The precision for 10 replicate determinations at the 5 ng mL-1 Pb levels was 2.8% relative standard deviation (RSD), calculated with the peak heights obtained. The calibration graph using the preconcentration system for lead was linear with a correlation coefficient of 0.9997 at levels near the detection limits up to at least 100 ng mL-1. The method was successfully applied to the determination of lead in drinking water samples.

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa